Multi-electrode array with a planar surface for cell patterning by microprinting

Loading...
Thumbnail Image

Authors

Slavík, Jan
Skopalík, Josef
Provazník, Valentýna
Hubálek, Jaromír

Advisor

Referee

Mark

Journal Title

Journal ISSN

Volume Title

Publisher

MDPI
Altmetrics

Abstract

Multielectrode arrays (MEAs) are devices for non-invasive electrophysiological measurements of cell populations. This paper describes a novel fabrication method of MEAs with a fully planar surface. The surface of the insulation layer and the surface of the electrodes were on one plane; we named this device the planar MEA (pMEA). The main advantage of the pMEA is that it allows uniform contact between the pMEA surface and a substrate for positioning of microfluidic channels or microprinting of a cell adhesive layer. The fabrication of the pMEA is based on a low adhesive Au sacrificial peel-off layer. In divergence from conventional MEAs with recessed electrodes, the electrodes of the pMEA lead across the sloped edge of the insulation layer. To make this, the profile of the edge of the insulation layer was measured and the impedance of the planar electrodes was characterized. The impedance of the pMEA was comparable with the impedance of conventional MEA electrodes. The pMEA was tested for patterning HL-1 cells with a combination of imprinting fibronectin and coating by antifouling poly (l-lysine)-graft-poly(ethylene glycol) (PLL-g-PEG). The HL-1 cells remained patterned even at full confluency and presented spontaneous and synchronous beating activity.
Multielectrode arrays (MEAs) are devices for non-invasive electrophysiological measurements of cell populations. This paper describes a novel fabrication method of MEAs with a fully planar surface. The surface of the insulation layer and the surface of the electrodes were on one plane; we named this device the planar MEA (pMEA). The main advantage of the pMEA is that it allows uniform contact between the pMEA surface and a substrate for positioning of microfluidic channels or microprinting of a cell adhesive layer. The fabrication of the pMEA is based on a low adhesive Au sacrificial peel-off layer. In divergence from conventional MEAs with recessed electrodes, the electrodes of the pMEA lead across the sloped edge of the insulation layer. To make this, the profile of the edge of the insulation layer was measured and the impedance of the planar electrodes was characterized. The impedance of the pMEA was comparable with the impedance of conventional MEA electrodes. The pMEA was tested for patterning HL-1 cells with a combination of imprinting fibronectin and coating by antifouling poly (l-lysine)-graft-poly(ethylene glycol) (PLL-g-PEG). The HL-1 cells remained patterned even at full confluency and presented spontaneous and synchronous beating activity.

Description

Citation

SENSORS. 2019, vol. 19, issue 24, p. 1-11.
https://www.mdpi.com/1424-8220/19/24/5379

Document type

Peer-reviewed

Document version

Published version

Date of access to the full text

Language of document

en

Study field

Comittee

Date of acceptance

Defence

Result of defence

Endorsement

Review

Supplemented By

Referenced By

Creative Commons license

Except where otherwised noted, this item's license is described as Creative Commons Attribution 4.0 International
Citace PRO