Impact of gain on high-frequency contrast-enhanced ultrasound video and radio frequency data

dc.contributor.authorŠikner, Tomáš
dc.contributor.authorKolář, Radim
dc.coverage.issue3cs
dc.coverage.volume19cs
dc.date.accessioned2023-10-09T08:32:18Z
dc.date.available2023-10-09T08:32:18Z
dc.date.issued2017-06-30cs
dc.description.abstractWe have examined the impact of a gain on contrast-enhanced ultrasound video (JPEG DICOM) and radiofrequency (RF) data. The gain is one of main acquisition parameters and therefore, its influence on a data analysis might be huge. The gain was set in a range of 10 to 50 dB. Imaging was performed with an ultrasound system Vevo 2100 and a linear array probe MS 250 with a transmit frequency of 18 MHz. A contrast agent used in this study was Sonovue (Bracco, Milan, Italy). Measurements were performed for two solutions with a different contrast agent concentration (a dilution factor of 1:4 and 1:16, a solvent is saline). The impact of the gain was evaluated using three parameters: the contrast to background ratio (CBR), the contrast to noise ratio (CNR) and the background to noise ratio (BNR). In RF data, an image intensity grows up continually with the increasing gain but in video data, saturation can occur at higher values of the gain. In terms of the CBR, CNR and BNR, the impact of the gain on RF data is negligible but on video data is substantial.en
dc.formattextcs
dc.format.extent73-77cs
dc.format.mimetypeapplication/pdfen
dc.identifier.citationElektrorevue. 2017, vol. 19, č. 3, s. 73-77. ISSN 1213-1539cs
dc.identifier.issn1213-1539
dc.identifier.urihttp://hdl.handle.net/11012/214235
dc.language.isocscs
dc.publisherInternational Society for Science and Engineering, o.s.cs
dc.relation.ispartofElektrorevuecs
dc.relation.urihttp://www.elektrorevue.cz/cs
dc.rights(C) 2017 Elektrorevueen
dc.rights.accessopenAccessen
dc.titleImpact of gain on high-frequency contrast-enhanced ultrasound video and radio frequency datacs
dc.title.alternativeImpact of gain on high-frequency contrast-enhanced ultrasound video and radio frequency dataen
dc.type.driverarticleen
dc.type.statusPeer-revieweden
dc.type.versionpublishedVersionen
eprints.affiliatedInstitution.facultyFakulta elektrotechniky a komunikačních technologiícs