Comparing Saddle, Slotted-tube and Parallel-plate Coils for Magnetic Resonance Imaging
dc.contributor.author | Nešpor, Dušan | cs |
dc.contributor.author | Bartušek, Karel | cs |
dc.contributor.author | Dokoupil, Zdeněk | cs |
dc.coverage.issue | 3 | cs |
dc.coverage.volume | 14 | cs |
dc.date.issued | 2014-12-01 | cs |
dc.description.abstract | The paper is concerned with a comparison of the properties of RF coils of three configurations for MRI measurements on small animals. In comparison with the classical saddle coil the proposed modification of slotted-tube coil exhibits identical homogeneity of B1 field in a larger space. The parallel-plate coil has a satisfactory homogeneity of B1 field over the whole internal space. The signal-to-noise ratio measured for all three coils is roughly the same and is given by the magnitude of RF pre-amplifier noise. As the slotted-tube and parallel-plate coils have a lower inductance compared with the saddle coil, they can be tuned to resonance on the 200 MHz frequency even at larger dimensions. The results show that the parallel-plate coil has very good properties for the measurement of small animals. | en |
dc.description.abstract | Článe se zabývá porovnáním vlastností radiofrekvenčních cívek tří typů pro tomografické zobrazování malých zvířat magnetickou rezonancí. | cs |
dc.format | text | cs |
dc.format.extent | 171-176 | cs |
dc.format.mimetype | application/pdf | cs |
dc.identifier.citation | Measurement Science Review. 2014, vol. 14, issue 3, p. 171-176. | en |
dc.identifier.doi | 10.2478/msr-2014-0023 | cs |
dc.identifier.issn | 1335-8871 | cs |
dc.identifier.orcid | 0000-0002-3858-9434 | cs |
dc.identifier.orcid | 0000-0002-6598-5424 | cs |
dc.identifier.other | 111197 | cs |
dc.identifier.researcherid | D-3389-2012 | cs |
dc.identifier.scopus | 6508372019 | cs |
dc.identifier.uri | http://hdl.handle.net/11012/201757 | |
dc.language.iso | en | cs |
dc.publisher | De Gruyter Open | cs |
dc.relation.ispartof | Measurement Science Review | cs |
dc.relation.uri | https://www.sciendo.com/article/10.2478/msr-2014-0023 | cs |
dc.rights | Creative Commons Attribution-NonCommercial-NoDerivatives 4.0 International | cs |
dc.rights.access | openAccess | cs |
dc.rights.sherpa | http://www.sherpa.ac.uk/romeo/issn/1335-8871/ | cs |
dc.rights.uri | http://creativecommons.org/licenses/by-nc-nd/4.0/ | cs |
dc.subject | MRI measurements | en |
dc.subject | RF coils | en |
dc.subject | homogeneity of B1 | en |
dc.subject | numerical analysis | en |
dc.subject | radiofrekvenční cívka | |
dc.subject | homogenita magnetického pole | |
dc.subject | numerická analýza B1 | |
dc.subject | magnetická rezonance | |
dc.subject | tomografie | |
dc.title | Comparing Saddle, Slotted-tube and Parallel-plate Coils for Magnetic Resonance Imaging | en |
dc.title.alternative | Porovnání cívek typy Saddle, Slotted-tube and Parallel-plate Coils pro zobrazování magnetickou rezonancí | cs |
dc.type.driver | article | en |
dc.type.status | Peer-reviewed | en |
dc.type.version | publishedVersion | en |
sync.item.dbid | VAV-111197 | en |
sync.item.dbtype | VAV | en |
sync.item.insts | 2025.02.03 15:42:49 | en |
sync.item.modts | 2025.01.17 16:49:45 | en |
thesis.grantor | Vysoké učení technické v Brně. Fakulta elektrotechniky a komunikačních technologií. Ústav teoretické a experimentální elektrotechniky | cs |
Files
Original bundle
1 - 1 of 1
Loading...
- Name:
- 10.2478_msr20140023.pdf
- Size:
- 2.77 MB
- Format:
- Adobe Portable Document Format
- Description:
- 10.2478_msr20140023.pdf