Iterative Precise Conductivity Measurement with IDEs
dc.contributor.author | Hubálek, Jaromír | cs |
dc.coverage.issue | 5 | cs |
dc.coverage.volume | 15 | cs |
dc.date.issued | 2015-05-22 | cs |
dc.description.abstract | The paper presents a new approach in the field of precise electrolytic conductivity measurements with planar thin- and thick-film electrodes. This novel measuring method was developed for measurement with comb-like electrodes called interdigitated electrodes (IDEs). Correction characteristics over a wide range of specific conductivities were determined from an interface impedance characterization of the thick-film IDEs. The local maximum of the capacitive part of the interface impedance is used for corrections to get linear responses. The measuring frequency was determined at a wide range of measured conductivity. An iteration mode of measurements was suggested to precisely measure the conductivity at the right frequency in order to achieve a highly accurate response. The method takes precise conductivity measurements in concentration ranges from 106 to 1 M without electrode cell replacement. | en |
dc.description.abstract | Příspěvek prezentuje nový přístup v oblasti přesných elektrolytických měření vodivosti s planární tenkého a a tlustovrstvých elektrod. Tato nová metoda měření byl vyvinut pro měření s hřebenovitých elektrod zvané vzájemně do sebe zasahujících elektrod (IDE). Korekční charakteristiky v širokém spektru specifické vodivosti byly stanoveny z charakterizace impedanční rozhraní tlustovrstvých IDE. Používá se lokální maximum kapacitního části impedance rozhraní pro opravy, aby lineární odezvy. Měření frekvence byla stanovena při širokém rozsahu měřeného vodivosti. Iterace způsob měření, bylo navrženo, aby přesně měřit vodivost na správnou frekvenci, aby bylo dosaženo vysoce přesnou odezvu. Metoda trvá přesná měření vodivosti v koncentračních rozmezích od 10-6 do 1 m bez náhrady elektroda buněk. (strojový překlad) | cs |
dc.format | text | cs |
dc.format.extent | 12080-12091 | cs |
dc.format.mimetype | application/pdf | cs |
dc.identifier.citation | SENSORS. 2015, vol. 15, issue 5, p. 12080-12091. | en |
dc.identifier.doi | 10.3390/s150512080 | cs |
dc.identifier.issn | 1424-8220 | cs |
dc.identifier.orcid | 0000-0002-7496-2558 | cs |
dc.identifier.other | 118659 | cs |
dc.identifier.researcherid | D-7753-2012 | cs |
dc.identifier.scopus | 57207501711 | cs |
dc.identifier.uri | http://hdl.handle.net/11012/137230 | |
dc.language.iso | en | cs |
dc.publisher | MDPI | cs |
dc.relation.ispartof | SENSORS | cs |
dc.relation.uri | http://www.mdpi.com/1424-8220/15/5/12080 | cs |
dc.rights | Creative Commons Attribution 4.0 International | cs |
dc.rights.access | openAccess | cs |
dc.rights.sherpa | http://www.sherpa.ac.uk/romeo/issn/1424-8220/ | cs |
dc.rights.uri | http://creativecommons.org/licenses/by/4.0/ | cs |
dc.subject | IDEs | en |
dc.subject | electrolytic conductance | en |
dc.subject | specific conductivity | en |
dc.subject | iterative measurement | en |
dc.subject | cell constant correction | en |
dc.subject | IDE | |
dc.subject | elektrolytická vodivost | |
dc.subject | měrná vodivost | |
dc.subject | iterativní měření | |
dc.subject | korekce článkové konstanty | |
dc.title | Iterative Precise Conductivity Measurement with IDEs | en |
dc.title.alternative | Iterativní přesné měření vodivosti s IDE | cs |
dc.type.driver | article | en |
dc.type.status | Peer-reviewed | en |
dc.type.version | publishedVersion | en |
sync.item.dbid | VAV-118659 | en |
sync.item.dbtype | VAV | en |
sync.item.insts | 2025.02.03 15:40:53 | en |
sync.item.modts | 2025.01.17 16:50:19 | en |
thesis.grantor | Vysoké učení technické v Brně. Fakulta elektrotechniky a komunikačních technologií. Ústav mikroelektroniky | cs |