Calibrating the Six-Port Reflectometer Using a Matched Load And Four Unity-Reflection Standards (Part 1)

Loading...
Thumbnail Image
Date
1992-10
ORCID
Advisor
Referee
Mark
Journal Title
Journal ISSN
Volume Title
Publisher
Společnost pro radioelektronické inženýrství
Abstract
An alternative six-port calibration procedure is described using an ideal matched load and four unity reflection standards. Not only offset short or open circuits can be used as the reflection standards but also arbitrary reactances, such as inductors or capacitors. A different mathematical approach is employed to obtain the calibration equations.
Description
Citation
Radioengineering. 1992, vol. 1, č. 1, s. 10-13. ISSN 1210-2512
http://www.radioeng.cz/fulltexts/1992/92_01_02.pdf
Document type
Peer-reviewed
Document version
Published version
Date of access to the full text
Language of document
en
Study field
Comittee
Date of acceptance
Defence
Result of defence
Document licence
Creative Commons Attribution 3.0 Unported License
http://creativecommons.org/licenses/by/3.0/
DOI
Collections
Citace PRO