Measured Parameters Of The Three-Axis Gaussmeter

but.event.date25.04.2019cs
but.event.titleStudent EEICT 2019cs
dc.contributor.authorHejtmánek, Tomáš
dc.date.accessioned2020-04-16T07:19:40Z
dc.date.available2020-04-16T07:19:40Z
dc.date.issued2019cs
dc.description.abstractThe purpose of this article is a measuring parameters of a gaussmeter with a three-axis measuring probe designed for magnetic field mapping. Commercially available gaussmeters have low frequency bandwidth for these purposes, and signals cannot be measured in synchronous detection mode. Therefore, a suitable magnetic field sensor was selected and one channel of the device was designed and made. In particular, the higher bandwidth for three-axis measurement and the possibility of synchronous signal detection were taken into account. Attention will also be paid to designing a suitable mechanical design of the probe and the location of the sensors. The gaussmeter will be used for magnetic impedance tomography (MIT), which is reconstruction method of conductivity image. On the proposed device was performed measurements of frequency bandwidth, noise measuring of Hall probe and measuring constants of Hall generators. The properties were evaluated and compared with parameters of commercially available Gaussmeters.en
dc.formattextcs
dc.format.extent718-722cs
dc.format.mimetypeapplication/pdfen
dc.identifier.citationProceedings of the 25st Conference STUDENT EEICT 2019. s. 718-722. ISBN 978-80-214-5735-5cs
dc.identifier.isbn978-80-214-5735-5
dc.identifier.urihttp://hdl.handle.net/11012/186766
dc.language.isoencs
dc.publisherVysoké učení technické v Brně, Fakulta elektrotechniky a komunikačních technologiícs
dc.relation.ispartofProceedings of the 25st Conference STUDENT EEICT 2019en
dc.relation.urihttp://www.feec.vutbr.cz/EEICT/cs
dc.rights© Vysoké učení technické v Brně, Fakulta elektrotechniky a komunikačních technologiícs
dc.rights.accessopenAccessen
dc.subjectGaussmeteren
dc.subjectfrequency bandwidthen
dc.subjectHall probe noise measuringen
dc.subjectconstants of Hall cellsen
dc.titleMeasured Parameters Of The Three-Axis Gaussmeteren
dc.type.driverconferenceObjecten
dc.type.statusPeer-revieweden
dc.type.versionpublishedVersionen
eprints.affiliatedInstitution.departmentFakulta elektrotechniky a komunikačních technologiícs
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
718_eeict2019.pdf
Size:
890.29 KB
Format:
Adobe Portable Document Format
Description: