This article compares eight types of CO2 sensors using NDIR technology for accuracy, response time, and acquisition costs. The experiment included static and dynamic tests in a chamber and a duct. The Testo 935 and Rotronic CL11/CP11 sensors showed high accuracy and fast response, while the Winsen MH-Z16 offered the best performance-to-cost ratio. Other tested sensors – NetAtmo, Comet S3532, LaskaKit SCD41 and MH-Z14 exhibited inaccuracies, making them unsuitable for scientific applications.
Description
Citation
Juniorstav 2025: Selected Papers. Proceedings 27th International Scientific Conference Of Civil Engineering, s. 1-11. ISBN .