Abstract
A convenient and exact method of using a sliding load to the calibration of the single six-port reflectometer is described. Neither reflection coefficient nor positions of the sliding load settings need be known. Reflection coefficient of the load is not limited to small values. No numerical iterations are involved in the calibration.