Calibrating the Six-Port Reflectometer Using a Matched Load And Four Unity-Reflection Standards (Part 2)
Loading...
Files
Date
Authors
Bilik, V.
Advisor
Referee
Mark
Journal Title
Journal ISSN
Volume Title
Publisher
Společnost pro radioelektronické inženýrství
ORCID
Abstract
A convenient and exact method of using a sliding load to the calibration of the single six-port reflectometer is described. Neither reflection coefficient nor positions of the sliding load settings need be known. Reflection coefficient of the load is not limited to small values. No numerical iterations are involved in the calibration.
Description
Keywords
Citation
Radioengineering. 1992, vol. 1, č. 2, s. 2-8. ISSN 1210-2512
http://www.radioeng.cz/fulltexts/1992/92_02_01.pdf
http://www.radioeng.cz/fulltexts/1992/92_02_01.pdf
Document type
Peer-reviewed
Document version
Published version
Date of access to the full text
Language of document
en
Study field
Comittee
Date of acceptance
Defence
Result of defence
DOI
Collections
Endorsement
Review
Supplemented By
Referenced By
Creative Commons license
Except where otherwised noted, this item's license is described as Creative Commons Attribution 3.0 Unported License

